20
DMC Repair Manual
Parts Replacement
Rear Panel and XSYNC Assembly (Figure 2-3)
Removal
1. Remove the bottom extrusion.
2. Remove the two (2) screws that secures SCSI connector J5 to the rear panel.
3. Remove the two (2) screws securing the rear panel to the top extrusion.
4. Gently pull off the rear panel from the top extrusion being careful not to break the XSYNC
soldered cable leads from the XSYNC assembly.
5. Disconnect the XSYNC cable leads from connector J4 on the back end board.
Figure 2-3. Removing rear panel and XSYNC assembly
Installation
1. Reconnect the XSYNC cable to connector J4 on the back end board.
2. Position the rear panel onto the top extrusion being careful not to pinch or grimp the
XSYNC cable leads between the top extrusion and the rear panel.
3. Install and tighten the two (2) screws that secures the rear panel to the top extrusion.
4. Install and tighten the two (2) screws that secures SCSI connector J5 to the rear panel.
5. Install the bottom extrusion.
SCSI Connector
(J4)
Top Extrusion
Rear Panel
XSYNC Assembly
XSYNC Cable