50
DMC Repair Manual
Testing & Adjustment
Flair & Bloom
• Click on the Flair & Bloom button. A window appears on the Pretest VI panel indicating
that the Flair and Bloom test is running. It also prompts you place a piece of white
paper approximately 1 foot in front of the camera to check for dark spots. this test
also checks foil for flair bloom.
• Once white paper is placed in front of the camera, click on the OK button. A white
screen appears. Check for dark spots.
Shutter Light Leak
• Turn on Lamp.
• Click on the Shutter Light leak button. The Shutter Alignment window appears.
Also, a dark image of the target appears with a bright red LED in its middle.
• Turn off Lamp.
Summary of Contents for Digital Microscope Camera
Page 5: ...1 Camera Overview...
Page 13: ...2 Parts Replacement...
Page 37: ...3 Testing and Adjustment...