29
DMC Repair Manual
Parts Replacement
8. Place the backing pad over the CCD imager PC board making sure that the small and large
tabs are properly positioned.
9. Place the CCD retainer over the CCD imager and backing pad assembly.
10. Install and tighten the three (3) screws that secures the optics assembly to its front panel
supports.
11. Install the front panel and optics assembly onto the top extrusion making sure that the shutter
flex and the CCD imager flex are properly secured to their respective connector on the front
end PC board.
12. Install the bottom extrusion.