DMC Repair Manual
Parts Replacement
Installation
1. Reconnect the shutter flex to connector J1 and the CCD imager flex to connector J5 and on
the front end board.
2. Push in connector J1 clamp to secure the shutter flex and connector J5 to secure the
CCD imager flex.
3. Position the front panel onto the top extrusion being careful not to pinch or grimp the
shutter or CCD imager flex between the top extrusion and the front panel.
Note:
Make sure that the vanity panel is inserted in the front panel groove.
4. Install and tighten the two (2) screws that secures the front panel to the top extrusion.
5. Install the bottom extrusion.