29.9.3 Continuous cleaning
About this task
The disinfectant solution is mixed in the water container: A small amount of
Planmeca PlanClear disinfectant concentrate is regularly dosed to the water
container where it is mixed with water and fed further to the waterlines, the
instruments and the cup fill. For information on the disinfection
concentration, see section "Technical specifications" on page 273.
When continuous cleaning is enabled (factory default), it is shown as a small
green icon at the top of the control panel. When continuous cleaning is
disabled, the icon is grey.
NOTE
Even if continuous cleaning is enabled, you must still run the cleaning
programs as described in section "Flushing and cleaning programs" on page
NOTE
When you disable continuous cleaning, a message is displayed that advises
you to perform long flushing to remove excess disinfectant solution from the
waterlines. However, negligence to perform long flushing does not prevent
dental unit operation.
To disable/enable the continuous cleaning, follow the steps below:
Steps
1. Press Program.
29 Flushing and cleaning programs
User's manual
Planmeca Sovereign Classic 227
Summary of Contents for Sovereign Classic
Page 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Page 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Page 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Page 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Page 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Page 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Page 296: ......
Page 297: ......