29 Flushing and cleaning programs
29.1 When to use cleaning programs
We recommend that you run the flushing and cleaning programs as follows.
Recommended cleaning intervals
Icon
Cleaning program
When
Duration
Short flushing
After every patient
30 seconds / instrument
Long flushing
In the morning and after the
working day
2 minutes (default; can be
programmed)
Suction flushing
In the morning
< 3 minutes
Suction cleaning
After the working day
< 5 minutes
Waterline cleaning
Continuous
cleaning
enabled
When
needed. We
recommend
that waterline
cleaning be
performed
once a month
after the
working day
(but not over
the weekend).
Cleans the dental unit’s
waterlines.
The amount of disinfectant
concentrate used in a
cleaning cycle depends on
whether continuous cleaning
is enabled or disabled. For
more information, see table
Water and waterline
disinfection in section
Duration: < 10 minutes in the
e min. 8 hours affect
time + < 30 minutes in the
morning
Continuous
cleaning
disabled
When
needed. We
recommend
that waterline
cleaning be
performed
once a week
after the
working day
(but not over
the weekend).
29 Flushing and cleaning programs
User's manual
Planmeca Sovereign Classic 209
Summary of Contents for Sovereign Classic
Page 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Page 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Page 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Page 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Page 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Page 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Page 296: ......
Page 297: ......