CAUTION
If the drain is blocked, the dental unit might overflow with contaminated
water and excessive water could flow onto the floor. Shut down the unit and
contact your property's caretaker and your Planmeca dealer.
CAUTION
When a water leak is detected, help message HE4004 is displayed. Turn off
the clinic’s main water tap and contact your Planmeca dealer.
CAUTION
For surgical procedures, use sterile irrigating solutions, such as sterile water
or saline. Appropriate delivery devices should be used to deliver sterile
irrigating solutions during surgery. This may include a dedicated surgical
irrigation system with components including handpieces that are single-use
disposable or compatible with heat sterilisation methods used in outpatient
dental settings.
CAUTION
A power cut will shut down the software-controlled backflow prevention
system. If you are using a turbine without a built-in backflow prevention
system, contaminated water can enter into the turbine and turbine hose in
the event of a power cut.
CAUTION
In extreme operating conditions the temperature of the chair lift and backrest
motors might rise significantly. DO NOT TOUCH THE MOTORS!
CAUTION
In extreme operating conditions the surface temperature of the seat
upholstery may rise to 44°C (111°F). When applying maximum allowed load
on the patient chair in hot climate conditions, ensure sufficient time for the
seat upholstery to cool down.
NOTE
Before switching on the dental unit, make sure that the main water feed, air
pressure and suction motor are turned on.
NOTE
National regulations concerning the quality of dental water and dental air
must be followed when using the Planmeca dental unit.
NOTE
The water used by the dental unit instruments and cup fill are to be used for
rinsing only. For more information, please contact your Planmeca dealer.
NOTE
The main water feed must be turned off when the dental unit is not in use.
NOTE
The user must monitor the microbial load of the water used by the dental
unit. Biotest Plus (by Alpro Medical GmbH) or equivalent test can be used
for this purpose.
7 For your safety
User's manual
Planmeca Sovereign Classic 13
Summary of Contents for Sovereign Classic
Page 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Page 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Page 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Page 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Page 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Page 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Page 296: ......
Page 297: ......