If you are using the Morita TORX micromotor and the CA-10RC-ENDO
10:1 handpiece with the apex locator, touch the file with the contrary
electrode. Check that all the meter indicator bars in the
Apex locator
window light up.
If the indicator bars do not appear normally, stop using the device and
contact your Planmeca dealer.
24.3.2 Using apex locator
The Morita Root ZX mini U apex locator can be used both with hand files
and with the Morita TORX micromotor and the CA-10RC-ENDO 10:1
handpiece.
If you are using hand files, press Apex locator to open the
Apex locator
window.
NOTE
If the Apex locator button is not visible in your control panel's swipe menu,
you must add it. See section "Organising items on control panel" on page
If you are using the Morita TORX micromotor and the CA-10RC-ENDO 10:1
handpiece, the
Apex locator window automatically opens on the control
panel when you select one of the apical presets a1 - a3.
24 Operating instruments
User's manual
Planmeca Sovereign Classic 155