The pop-up window closes and the new values are displayed on the
control panel.
26.6 Timer settings
About this task
Up to six timer settings can be programmed into the timer memory.
When you have signed in to the dental unit with your own user name, all
changes you make to the timer settings and then save, are saved to your
user profile. This means that whenever you sign in to the dental unit, you will
use your last saved timers.
Steps
1. Press Program.
2. Press Timer.
A pop-up window opens.
3. Press the timer that you want to adjust.
A new pop-up window opens.
4. Adjust the length of the timer.
The minimum value is 5 sec. and the maximum 20 min. The adjustment
step is 5 sec.
5. Press OK.
The pop-up window closes.
26.7 Bowl rinse and cup fill
26.7.1 Adjusting cup fill and bowl rinse flow rates
The flow rates of the bowl flush and cup fill can be adjusted with the two
black knobs located inside the dental unit.
26 Programming
188 Planmeca Sovereign Classic
User's manual