If the flash bar has been set at 2 on the meter scale, the only signal that is
heard is the continuous signal when the file tip reaches the flash bar.
1. Continuous signal
24.3.3 Testing apex locator function
About this task
The apex locator function must be tested weekly.
NOTE
While the test is going on, the control panel is locked.
Steps
1. Open the
Apex locator window either by pressing the Apex locator button
or by activating the Morita TORX micromotor and selecting one of the
apex presets (a1 - a3).
2. Insert the tester into the jack on the back of the instrument console.
24 Operating instruments
158 Planmeca Sovereign Classic
User's manual