1. Switch off the dental unit.
2. Detach the sensor and leave it hanging next to the container.
3. Remove the cap, including the hose attached to it, from the container.
Disinfect the cap and hose in a washer-disinfector.
4. Carefully lift the water container upward to remove it from the cuspidor.
Disinfect the container in a washer-disinfector.
CAUTION
Be careful not to splash the contents of the water container when removing
it, as it may contain traces of Planmeca PlanClear disinfectant. Immediately
remove any Planmeca PlanClear splashes from the dental unit.
When you place the water container back in the cuspidor, make sure to
place the three adjacent openings in the container carefully over the three
pipes in the cuspidor. Attach the sensor and make sure you place the cap
back properly so that it is tightly closed.
30.5.3 Instrument flushing holder
Disinfect the instrument flushing holder once a month in a washer-disinfector
at 93°C. If needed, autoclave it at 134°C.
To remove the instrument flushing holder, follow the steps below:
30 Cleaning and disinfection
244 Planmeca Sovereign Classic
User's manual
Summary of Contents for Sovereign Classic
Page 1: ...PlanmecaSovereign Classic user s manual EN 10032652...
Page 9: ...35 7 RF Exposure 285 Table of contents User s manual Planmeca Sovereign Classic...
Page 10: ...Table of contents Planmeca Sovereign Classic User s manual...
Page 27: ...7 For your safety User s manual Planmeca Sovereign Classic 17...
Page 291: ...34 2 6 Top view 34 Technical information User s manual Planmeca Sovereign Classic 281...
Page 292: ...34 Technical information 282 Planmeca Sovereign Classic User s manual...
Page 296: ......
Page 297: ......