8 GCS Commands
136
Version: 1.2.1
MS249E
C-863.12 Mercury Controller
name of your positioner with SPA (p. 190) or SEP (p. 186).
You can find details in the parameter overview (p. 239).
CSV? (Get Current Syntax Version)
Description:
Queries the GCS syntax version used in the firmware.
Format:
CSV?
Arguments:
None
Response:
The current GCS syntax version
Notes:
1.0 (for GCS 1.0) or 2.0 (for GCS 2.0) are possible
responses.
CTO (Set Configuration Of Trigger Output)
Description:
Configures the trigger output conditions for the specified
digital output line.
Format:
CTO {<TrigOutID> <CTOPam> <Value>}
Arguments:
<TrigOutID> is one digital output line of the controller, see
below for details.
<CTOPam> is the CTO parameter ID in decimal format, see
below for the available IDs.
<Value> is the value that the CTO parameter is set to, see
below.
Response:
None
Notes:
The trigger output conditions will become active when
enabled with TRO (p. 200). Do not use DIO (p. 144) on
digital output lines where the trigger output is activated by
TRO.
The CTO settings are lost when you power down or reboot
the C-863.12. They can be easily maintained by saving
them in a macro.