6 Startup
C-863.12 Mercury Controller
MS249E
Version: 1.2.1
53
INFORMATION
A unique address must be set for each controller in a daisy chain network. One of the
controllers must have the address 1. This controller does not have to be the one directly
connected to the PC.
INFORMATION
A non-networked controller must have address 1, if it
▪
is to be used in PIMikroMove.
▪
is to be used in drivers for NI LabVIEW software.
▪
is to be addressed with PITerminal without specifying the target address; in this case, the
target and sender address (p. 117) are also not specified in the responses from the C-
863.12.
6.2.3
Baud Rate
Baud rate*
S5
S6
9600
ON
ON
19200
ON
OFF
38400
OFF
ON
115200
OFF
OFF
*Factory default settings are shown in bold.
INFORMATION
The same baud rate must be set for all controllers in a daisy chain network.