10 Maintenance
252
Version: 1.2.1
MS249E
C-863.12 Mercury Controller
7.
Close the
PI Firmware Updater
program by clicking the cross in the top right corner of
the window.
8.
Switch off the C-863.12 by pulling out the power cord of the power adapter from the
power socket.
9.
Switch the C-863.12 on again by connecting the power cord of the power adapter to the
power socket.
If the firmware update was successful, the C-863.12 exits the firmware update mode
and the
STA
LED lights up continuously.
Have new parameters been added by the firmware update, or has the memory management of
the C-863.12 been changed?
▪
If no: Firmware update is finished.
▪
If yes: Initialization of the C-863.12 is required, see below.
Initializing the C-863.12 after a firmware update
The initialization of the C-863.12 resets
all
parameters to their factory settings and deletes all
controller macros. Consequently, parameter values and controller macros that are not saved
are lost during the initialization process.
1.
Make sure that the current parameter values and controller macros of the C-863.12
have been saved on the PC.
2.
On the PC, start PITerminal or PIMikroMove, connect to the C-863.12, and, if necessary,
open the window to send commands.
3.
Initialize theC-863.12, by sending the following commands one by one:
ZZZ 100 parameter
ZZZ 100 macros
After successful initialization, the controller issues a corresponding message.