10-2
Care and maintenance
Exterior and
touchscreen cleaning
CAUTION:
To prevent possible damage to the ventilator, use only those cleaning
and disinfecting agents listed in this manual.
CAUTION:
To prevent possible damage to the ventilator, do not drip or spray any
liquids directly onto any surface including the front panel,
touchscreen, navigation ring (legacy versions), and Accept button.
CAUTION:
CAUTION:
Do not attempt to sterilize or autoclave the ventilator.
NOTE:
Use of unapproved cleaning and disinfecting agents may cause
damage to the enclosure, touchscreen, or parts of the ventilator.
Approved cleaning agents
The following cleaning agent is acceptable for use on the touchscreen and
exterior surfaces of the ventilator:
•
Medivators Intercept Detergent, per manufacturer's recommendation
at 1/3 oz (10 mL) per gallon of warm tap water.
Cleaning instructions
1. Apply cleaning agent to a soft lint-free cloth or use a disposable wipe.
The cloth or wipe should be saturated but not dripping.
2. Wipe cleaning agent over the entire exterior surface and touchscreen
of the ventilator.
3. Continue wiping until all visible contaminants and soiling are
removed.
4. Rinse with a clean, water-dampened cloth and allow to dry completely
before reuse.
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......