D-2
Regulatory compliance
- Connect the equipment to an outlet on a different circuit from that to
which the other devices are connected and contact Philips Service for
help.
WARNING:
Take care when operating the ventilator around other equipment, to avoid
reciprocal interference. Potential electromagnetic interference (EMI),
electrostatic discharge (ESD), or other interference can occur to the
ventilator or other equipment. Interference can interrupt ventilator
operation or degrade system performance, which can result in patient
injury. Try to minimize this interference by not using other equipment in
conjunction with the ventilator. If adjacent or stacked use is necessary,
follow the recommendations for placement of the equipment in
“Electromagnetic compatibility declaration” on page D-1. Monitor the
ventilator and nearby devices to verify normal operation.
WARNING:
Do not use the ventilator near active high frequency (HF) surgical
equipment, medical devices such as X-ray devices and diathermy, or in
an RF-shielded room of medical equipment or system for magnetic
resonance imaging, where the intensity of electromagnetic (EM)
disturbances is high. These devices may degrade the performance of the
ventilator.
Electromagnetic emissions
NOTE:
The emissions characteristics of this equipment make it suitable for
use in industrial areas and hospitals (CISPR 11 class A).
Guidance and manufacturer's declaration - electromagnetic emissions
The V60/V60 Plus Ventilator is intended for use in the electromagnetic environment specified
below. The user of the V60/V60 Plus Ventilator should assure that it is used in such an envi-
ronment.
Emissions test
Compliance
Electromagnetic enforcement - guidance
RF Emissions CIS-
PR 11
Group 1
The V60/V60 Plus Ventilator uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause interference in nearby
electronic equipment.
RF Emissions CIS-
PR 11
Class A
The V60/V60 Plus Ventilator is suitable for use in all es-
tablishments, except for those that are domestic or di-
rectly connected to the public low voltage power supply
network that supplies buildings used for domestic pur-
poses.
Harmonic emis-
sions
IEC 61000-3-2
Class A
Voltage fluctua-
tions/flicker emis-
sions
IEC 61000-3-3
Complies
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......