Care and maintenance
10-5
Preventive
maintenance
WARNING:
WARNING:
The expected service life of your V60/V60 Plus ventilator is 10 years. Perform
preventive maintenance on the ventilator according to the schedule in
Table 10-2. You can view the hours of ventilator operation in the Vent Info
window (“Vent Info (ventilator information)” on page 6-18). The following
subsections provide details for some of these preventive maintenance
procedures.
Table 10-2: Schedule of preventive maintenance
Frequency
Component
Maintenance
Between patients
and per institu-
tional guidelines
Patient circuit
Per manufacturer recommendations.
Main flow bacteria
filter
Replace per institutional guidelines.
Every month
Cooling fan filter
Inspect for occlusions, dust, lint, etc. If discol-
ored or dirty, remove and wash or rinse thor-
oughly, and let dry completely before
reinstalling.
Air inlet filter
Inspect and replace if needed
Every year
Backup battery
Inspect, test, and replace if needed
*
* Must be done by authorized service personnel according to the instructions in the service
manual.
Ventilator
Preventive maintenance
*
As required
Backup battery
A new backup battery should be installed and
charged within one year of the date of manu-
facture identified on the battery and on the
shipping box.
Every 5 years
Backup battery
Replace.
*
Battery replacement is based on the
date of manufacture recorded on the battery
label. Also viewable in Diagnostic Mode on the
system information screen.
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......