4-4
Principles of operation
Shape signal method of cycling and triggering. The shape signal or “shadow
trigger” method uses a mathematical model derived from the flow signal. A
new flow signal (shape signal) is generated by offsetting the signal from the
actual flow and delaying it (Figure 4-2). This intentional delay causes the flow
shape signal to be slightly behind the patient’s flow signal. If there is a sudden
change in patient flow, the patient’s flow signal crosses the shape signal; this
results in a trigger or a cycle. As a result, a sudden decrease in expiratory flow
from an inspiratory effort will cross the shape signal and create a signal for
ventilator triggering.
Figure 4-2: Shape signal
SET method of cycling. Patient flow reaches the spontaneous exhalation
threshold (SET); see Figure 4-3. The SET represents the intersection of the
flow waveform and a line of a given slope. SET is updated each breath.
Figure 4-3: Spontaneous exhalation threshold (SET)
Estimated
patient flow
Shape
signal
Cycle to exhalation
crossover point
Trigger to
inspiration
crossover point
Spontaneous
exhalation threshold
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......