First-time installation
A-5
7. Attach the BATT label as shown in Figure 3-5 on page 3-9.
WARNING:
Never attempt to disconnect or connect the battery during operation.
CAUTION:
Following battery installation, if a Check Vent or Vent Inoperative
alarm occurs when verifying ventilator operation, discontinue use of
the ventilator immediately and contact Philips. The Vent Inoperative
alarm occurs if AC power is disconnected and a battery is not
installed, or if the battery is fully discharged.
NOTE:
A new battery must be charged for at least 5 hours before being
placed into service. Based on the age and state of the battery, it may
take up to 16 hours or more to fully charge the battery.