Principles of operation
4-9
PCV mode
The PCV (pressure-controlled ventilation) mode delivers pressure-controlled
breaths, either triggered by the ventilator (Timed) or the patient (Spont). The
control settings active in the PCV mode are shown in Figure 4-8. The IPAP
setting defines the applied inspiratory pressure for all breaths. If the patient
fails to trigger a breath through Auto-Trak within the interval determined by the
rate setting, the ventilator triggers a mandatory breath. The I-Time setting is
the cycle criterion for all breaths. Figure 4-9 shows a PCV mode pressure
waveform.
Figure 4-8: PCV controls
Figure 4-9: PCV pressure waveform
Pressure
EPAP
IPAP
I-Time
1/Rate
Machine-triggered
(Timed) breath
Rise
Time
Patient-triggered
(Spont) breath
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......