9-14
Alarms, messages, and troubleshooting
Check Vent: Blower
Stalled
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(2)
Yes
No
No
Check Vent: Blower
Temperature High
Technical failure
Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(33)
Yes
No
No
Check Vent: Cooling
Fan Speed Error
Overheating of ventila-
tor possible
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(36)
Yes
No
No
Check Vent: CPU
PCBA ADC Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(29)
Yes
No
No
Check Vent: Data Ac-
quisition PCBA ADC
Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(27)
Yes
No
No
Check Vent: Flash
File System Error
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(37)
Yes
No
No
Check Vent: Internal
Temperature High
CPU
Technical failure
Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(30)
Yes
No
No
Check Vent: Internal
Temperature High
Daq
Technical failure
Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(31)
Yes
No
No
Check Vent: Internal
Temperature High
Mtr
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(32)
Yes
No
No
Check Vent: Ma-
chine Pressure Sen-
sor Auto-Zero Failed
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(9)
Yes
No
No
Table 9-4: Check Vent alarm messages: summary and troubleshooting (continued)
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......