Care and maintenance
10-3
Exterior and
touchscreen
disinfection
CAUTION:
To prevent possible damage to the ventilator, use only those cleaning
and disinfecting agents listed in this manual.
CAUTION:
CAUTION:
CAUTION:
Do not attempt to sterilize or autoclave the ventilator.
NOTE:
Use of unapproved cleaning and disinfecting agents may cause
damage to the enclosure, touchscreen, or parts of the ventilator.
Approved disinfecting agents
The following disinfecting agents are acceptable for use on the touchscreen
and exterior surfaces of the ventilator:
Disinfection
instructions
1. Ensure that cleaning was done per the procedure in “Exterior and
touchscreen cleaning” on page 10-2.
2. Apply disinfecting agent to a soft lint-free cloth or use a disposable
wipe. The cloth or wipe should be saturated but not dripping.
3. Wipe disinfecting agent over the entire exterior surface of the
ventilator.
4. Allow disinfectant to remain on the surface for the following
contact times:
• bleach solution — 2 minutes
• hydrogen peroxide — 15 minutes
5. Rinse with a clean cloth dampened with water and allow to dry
completely before reuse.
Table 10-1: Exterior disinfection
Disinfectant
Solution of 1 part 5% sodium hypochlorite (bleach) diluted in 9 parts deionized water.
3% hydrogen peroxide
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......