6-26
Operation
Table 6-4 lists the settable alarms. (For a complete list of non-settable alarms,
including Patient Disconnect, Occlusion, Pressure Regulation High, and other
alarms, refer to Table 9-3 on page 9-7.)
Trigger
(optional)
Trigger Sensitivity. Auto-Trak+ employs several algorithms to determine the point at which
the inspiration begins. The larger the value, the more sensitive the trigger (that is, the pa-
tient can trigger inspiration with less effort). Normal is the Auto-Trak setting used when
Auto-Trak+ is not enabled.
Applies only when the optional Auto-Trak+ feature is installed.
Normal, +1 to
+7
V
T
(AVAPS Tar-
get Tidal Vol-
ume)
Target tidal volume to be delivered during inspiration. The ventilator meets this target by
adjusting the inspiratory pressure with each breath.
Applies in AVAPS mode only.
200 to
2000 mL
Table 6-3: Modes and control settings with ranges (continued)
Setting
Description
Range
Table 6-4: Alarm settings
Setting
Description
Range
Hi Rate (High Rate Alarm)
High total breath rate.
5 to 90 BPM
Lo Rate (Low Rate Alarm)
Low total breath rate.
1 to 89 BPM
NOTE:
In non-CPAP modes, the Low Rate Alarm is
essentially off if set below the Respiratory Rate
setting.
Hi V
T
(High Tidal Volume
Alarm)
High exhaled tidal volume.
200 to 3500 mL
Lo V
T
(Low Tidal Volume
Alarm)
Low exhaled tidal volume.
OFF to 1500 mL
HIP (High Inspiratory Pres-
sure Alarm)
High pressure at the patient
airway.
5 to 50 cmH
2
O
LIP (Low Inspiratory Pres-
sure Alarm)
Low pressure at the patient
airway.
OFF to 40 cmH
2
O
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......