Alarms, messages, and troubleshooting
9-7
Alarms and other
messages
Table 9-3 is a list of alarms and other messages displayed by the ventilator,
along with descriptions, suggested corrective actions, and other information.
The ID (identifier) listed with the priority type is the priority number of the
alarm. This priority number determines the order of alarm message display.
Unless otherwise indicated, alarms listed as autoresettable are reset when the
alarm condition is removed.
Table 9-3: Alarm and other messages: summary and troubleshooting
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
AVAPS: Target V
T
Ex-
ceeded. Min Pres-
sure Too High
AVAPS target pressure is
less than Min P setting.
The ventilator limits its
applied pressure to Min P.
Check the patient. Con-
firm pressure settings are
compatible with target.
Evaluate pressure and vol-
ume settings.
Infor-
mation
(63)
No
Yes
N/A
AVAPS: Target V
T
Not
Achieved. Insuffi-
cient Max Pressure
AVAPS target pressure ex-
ceeds Max P setting. The
ventilator limits applied
pressure to Max P.
Check the patient. Con-
firm pressure settings are
compatible with target.
Evaluate pressure and vol-
ume settings.
Infor-
mation
(62)
No
Yes
N/A
Bacteria filter must
be installed onto gas
outlet
An inspiratory bacteria fil-
ter must be installed on
the patient gas outlet port.
Confirm that a bacteria fil-
ter is installed. Install a
bacteria filter if one is not
present.
Infor-
mation
(67)
Yes
N/A
N/A
Cannot Reach Target
Flow
Displays when HFT (high
flow therapy) is active. In-
dicates that flow target is
not achieved.
Check the patient. Check
that the high flow nasal
cannula size is appropri-
ate for the flow setting.
Check that an occlusive
interface is NOT in use (a
cannula fully sealed with-
in the nares, an NIV mask
or direct connection to an
ETT/trach). Check for an
occlusion, kink or liquid in
the patient circuit.
Low
(53)
No
Yes
Yes
Check Vent:
description of failure
High Inspiratory Pres-
sure
Measured inspiratory pres-
sure is greater than the
HIP setting, and the venti-
lator cycles into exhala-
tion. Autoresets after a
complete inspiration with-
out the alarm condition.
Check the patient. Con-
firm ventilator and alarm
settings are appropriate. If
problem persists, provide
alternative ventilation.
Have the ventilator ser-
viced.
High
(45)
Yes
Yes
Yes
High O
2
Supply Pres-
sure
O
2
inlet pressure is great-
er than 92 psig, so O
2
en-
richment ends. Autoresets
when O
2
supply pressure
falls below 87 psig.
Check the patient. If prob-
lem persists, provide alter-
native ventilation. Have
the ventilator serviced.
High
(49)
No
Yes
Yes
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......