Setting up the ventilator for use
5-9
d. VERIFY that the blue “Running on Internal Battery” message is
displayed.
6. Install a fully-charged battery and reconnect the ventilator to AC
power.
7. Repeat step 5 to verify that the ventilator switches over to battery
power.
Running alarm tests
The ventilator performs a self-check during start-up and continuously during
operation. Alarm functionality is verified by this self-check. You may also want
to run alarm tests, which demonstrate the alarms’ operation. Follow these
steps to perform the tests.
WARNING:
Preparation
1. Set the ventilator up as for normal ventilation, complete with
breathing circuit (PN 582073 or the equivalent) and a 1-liter test lung
assembly (PN 1021671).
2. Set the mode to S/T and make the following control settings: Rate:
4 BPM, IPAP: 10 cmH
2
O, EPAP: 6 cmH
2
O, I-Time: 1 sec, Rise: 1,
Ramp: Off, O
2
: 21%.
3. Make the following alarm settings: Hi Rate: 90 BPM, Lo Rate: 1 BPM,
Hi V
T
: 2000 mL, Lo V
T:
OFF, HIP: 50 cmH
2
O, LIP: OFF, Lo
E
: OFF,
LIP T: 5 secs.
High Inspiratory Pressure
1. Lower the HIP alarm limit to 8 cmH
2
O.
2. VERIFY that the
High Inspiratory Pressure
alarm is activated, the ventilator
cycles into exhalation, and pressure falls to 6 cmH
2
O (the EPAP
level).
3. Raise the HIP alarm limit to 15 cmH
2
O.
V
•
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......