6-18
Operation
Mask/Port
See “Selecting the mask and exhalation port” on page 6-12.
Vent Info (ventilator information)
The Ventilator Information window displays software version and other
information specific to your ventilator.
Screen Lock
Screen Lock deactivates all buttons and tabs on the touchscreen except Alarm
Silence, Alarm Reset, the Alarm/Message button, and Help. Tabs are grayed
out as in this example.
This message bar is displayed at the top of the screen:
To unlock the screen, press the Accept button
on the top-right front of the
ventilator.
NOTE:
If Screen Lock is active, the touchscreen remains locked even if an
alarm becomes active.
Auto-Trak+
The Normal Auto-Trak settings work well for most patients. Pediatric patients,
however, may benefit from more sensitive trigger settings, while some adult
patients may benefit from more or less sensitive cycle settings.
Changing Auto-Trak+ settings
1. Select Auto-Trak+ from the Menu window.
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......