9-12
Alarms, messages, and troubleshooting
Proximal Pressure
Line Disconnect
Proximal pressure low for
a few seconds. Proximal
pressure line is discon-
nected. Air flow to the pa-
tient continues.
Check the patient. Recon-
nect proximal pressure
line. Confirm ventilator
and alarm settings are ap-
propriate. If problem per-
sists, provide alternative
ventilation. Have the ven-
tilator serviced.
High
(41)
Yes
Yes
Yes
Running on Internal
Battery
System is powered by the
internal battery. Autore-
sets when ventilator is
connected to AC power.
Connect ventilator to AC
power.
Low
(60)
Yes
Yes
Yes
Tx,
E-Cycle: +x
Use Menu to change
Auto-Trak+ is active and
using the displayed set-
tings. This messages is
displayed for 5 min after
start-up.
Confirm that Auto-Trak+
settings are appropriate.
Infor-
mation
(65)
Yes
Yes
N/A
Using Default Set-
tings
Displayed after power on
if setting values are cor-
rupted or not set, or if de-
fault values were restored
by the user.
Check the patient. Check
and adjust settings as re-
quired.
Infor-
mation
(61)
Yes
Yes
N/A
Vent Inoperative x
description of failure
Table 9-4: Check Vent alarm messages: summary and troubleshooting
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
Check Vent: 1.8 V
Supply Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(21)
Yes
No
No
Check Vent: 3.3 V
Supply Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(22)
Yes
No
No
Check Vent: 5 V Sup-
ply Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(23)
Yes
No
No
Check Vent: 12 V
Supply Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(24)
Yes
No
No
Check Vent: 24 V
Supply Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(25)
Yes
No
No
Table 9-3: Alarm and other messages: summary and troubleshooting (continued)
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......