5-12
Setting up the ventilator for use
Figure 5-4: Duration of cylinder oxygen (13,790 kPa/2000 psig) at various
leak rates
Leak of 50 L/min
a. V
T
= 500 mL, Rate = 40 BPM, EPAP = 6 cmH
2
O, IPAP = 18 cmH
2
O
Leak of 10 L/min
Leak of 20 L/min
Leak of 30 L/min
Leak of 40 L/min
FIO
2
(%)
0
10
20
30
40
50
60
70
80
90
Dura
ti
o
n
(mi
n
)
30
35
40
45
50
55
60
65
70
75
80
85
90
95 100
b. V
T
= 500 mL, Rate = 20 BPM, EPAP = 6 cmH
2
O, IPAP = 18 cmH
2
O
20
40
60
80
100
120
140
0
D
u
ration (mi
n
)
Leak of 10 L/min
Leak of 20 L/min
Leak of 30 L/min
Leak of 40 L/min
Leak of 50 L/min
30
35
40
45
50
55
60
65
70
75
80
85
90
95 100
FIO
2
(%)
Summary of Contents for Respironics V60
Page 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Page 2: ......
Page 8: ...viii ...
Page 28: ...2 10 Symbols ...
Page 44: ...3 16 General information ...
Page 60: ...4 16 Principles of operation ...
Page 74: ...5 14 Setting up the ventilator for use ...
Page 102: ...6 28 Operation ...
Page 110: ...7 8 High flow therapy ...
Page 114: ...8 4 Patient monitoring ...
Page 132: ...9 18 Alarms messages and troubleshooting ...
Page 142: ...10 10 Care and maintenance ...
Page 162: ...A 10 First time installation ...
Page 182: ...D 10 Regulatory compliance ...
Page 198: ...E 16 Diagnostic mode ...
Page 210: ...Index Index 6 ...
Page 211: ......