G 03910 - Edition 04- August 12
3/6
C 90
Use of the serial link (M4 monitoring)
List of controls
Order
Parameter
Response
Action
ADR
XXX
OK or ERR2
To allocate a specific adress to a product making up a network.
From 000 to 255 (can be modified only when the pump is
stopped).
DLI
XXX
OK ou ERR2
Program the automatic transmission interval in min:1 to 255 sec.
DLR
None
OK
Start up the Data Logger for RS232 only.
ECH
ON or OFF
OK or ERR2
Return all characters received on the serial link in the case of
ECHON for RS232 only.
HDR
<val> see folio 6/6
OK or ERR3
Change the response heading character.
Input its ASCII value.
IDN
None
V X.X
Send back the software version number.
SET
XXYZZZZ
OK or ERR2
Parameter setting: (example:
# 0 0 0 S E T X X Y Z Z Z Z
)
XX = 00 : FB power
01 :
LP
temperature
02 :
Analog
input
03 :
Nitrogen
04
: E1 Logic input
05
: E2 Logic input
06* : Purge flow prolonged
07 :
FB
maintenance
08 :
Reserved
09 :
Reserved
10 :
Reserved
11 :
Reserved
12 : Reserved
13 : Reserved/MFS alert
14 : Date setting : Day
15
: Date setting : Month
16
: Date setting : Year
17
: Time setting : Hour
18
: Time setting : Minute
19 :
Free
20
: Sel(0)/HP T° alarm (1)
21 :
Reserved
22
: E3 Logic input
Y = 0
: Alert (concerns parameters)
1
: Alarm XX=0/01/02/03/04/05/22)
ZZZZ =
Threshold value (see page 5/6)
SEP
<val> see folio 5/6
OK or ERR3
Used to change the separator character between 2 consecutive
data items in a response. Input its ASCII value.
SYS
ON or OFF
OK or ERR2
Start up the pumping unit with SYSON and stop with SYSOFF.
TPS
XX
Maintenance time
XX = 00 : Running time
01 : Full overhaul
02 : Reserved
03 : Reserved
04 : Reserved
05 : Reserved
06 : Reserved
* When XX=06 then
Y=0 : purge flow prolonged for 10 minutes, setting values from 0 to 120 min.
Y=1 : standby purge for 10 seconds, setting values from 0 to 9999s.
adixen Vacuum Products - A3H Series Operating instructions