G 03915 - Edition 02 - July 10
1/2
D 400
Pump or Roots does not start
Mains presence light indicator does not switch on ..........
Hand-held remote control display is not lighting up .......
Display is showing a message or an alarm ........................
page 1
page 1
page 3
page 3
page 4
page 4
Incident
Cause
Troubleshooting
Mains presence
green indicator light
does not switch on
No mains
Check power supply.
•
Main switch on position “0”
Switch it on “
1
”.
•
External emergency stop engaged
Check the presence of cover plug
J1
, or check its locking.
•
Unit emergency stop engaged
Rotate to unlock the emergency
stop on the front panel display.
•
CB1 breaker switched off
Switch it on (
D 10
).
•
CB2 breaker switched off
Switch it on (
D 10
).
•
Other problem
Call customer service center.
Mains presence light indicator does not switch on
Incident
Cause
Troubleshooting
Hand-held remote is not
lighting up,
but the “mains presence“
green light indicator is on
CB3 breaker switched off
Switch it on (
D 10
).
•
HHR cable incorrectly connected
Check the connection.
•
Other problem
Call customer service center.
Hand-held remote is not lighting up
adixen Vacuum Products - A3H Series Operating instructions