G 03532 - Edition 03 - June 10
1/1
E 14
Disconnecting the pump
from the installation
Switch off the pump by positionning the main switch to “0”.
Switch off the customer main circuit breaker.
Disconnect the main plug or the main wires from the main
disconnect switch (see
B 50
).
Switch off
the pump
Study the safety instructions for pump relocation before disconnecting
the pump (see
E 11
).
Disconnecting
the pumping system
Disconnect all electrical connectors (J1, J14, RS232 and RS485...).
Do not forget to connect the ”REMOTE CONTROL” and
”EMERGENCY STOP” plugs (supplied with the pump).
Disconnect the water inlet and outlet quick connectors.
Disconnect the N
2
purge connector.
Disconnect the pump from the inlet piping and fit a blanking plate
on the inlet.
Disconnect the secondary exhaust port from the customer
application.
Disconnect the pump from the exhaust piping and fit a blanking
plate on the exhaust.
Adjust the levelling feet so that the pump rests on the castors.
Remove the pump from the installation (see
B 20
for inverse
procedure).
Before disconnecting the pump from the installation, be sure to have
at your disposal the connecting accessories to blank-off the inlet and
exhaust ports. Refer to
E 15
for connecting accessory part numbers.
Continue by Conditionning the pump for shipping ..............
E 15
E 15
adixen Vacuum Products - A3H Series Operating instructions