G 03910 - Edition 04- August 12
5/6
C 90
Use of the serial link (M4 monitoring)
Modifying the
separator character
Order
Parameter
Response
Action
SEP
<val>
val = decimal
value of the
ASCII code of
the separator
character between
1 and 255
OK or ERR3
Change the separator character between two consecutive data
items in the response.
Note:
- The factory configuration is the decimal code «032» of the
«space» character.
- Customer selection is memorized by the monitoring.
An order of the type:
#000SEP044<CR>«044»
is the decimal ASCII
code of the character «,» (comma).
Command
#000STA<CR>
Response
#000,00000,2,0080,2530,0127,1230,0000000000 <CR>
Only commands requiring the separator character are affected by
the SEP order.
Example:
Modifying the
header character
Order
Paramater
Response
Action
HDR
<val>
val = decimalv
alue of the ASCII
code of the
header character
between 001
and127
OK or ERR3
Change the header character in front of a response address.
Note :
- The factory configuration is the decimal code «035» of the
character «#».
- The code «020» corresponds to «no header character»
- Customer selection is memorized by the monitoring.
The value «047», the decimal ASCII code of the character «/».
Command
#002HDR047<CR>
Response
/0020K
Command
#002IDN<CR>
Response
/002M5_A1503H HV V1.07
Example:
adixen Vacuum Products - A3H Series Operating instructions