2/4
C 43
Use of the M4 monitoring system for
pumping operation
G 03903 - Edition 04 - August 12
Start up the pump by pressing
START
.
The following operations will perform automatically:
inlet valve opens (if preselected),
FB203 pump starts up,
data capture and sensor processing,
purge gas valve opens if option
CMD.PURGE
disabled
Roots pump starts up (if selected).
At the same time the operating parameters are displayed.
Gain access to other operating parameters* using
+
or
–
keys.
(*Only installed and configurated options parameters are
displayed.)
Note: for the thermal regulation of the pumps, the water valves
open when the temperature reaches the FB temperature set value.
They close when the temperature decreases - 2
°C below the FB203
temperature set value.
Pump start-up
The pumps are equipped with an exhaust pressure monitoring sensor.
It is the responsibility of the user to prevent the risks related to excess
overpressure.
52H LOC N2 ROOTS
FB POWER 1180 W
PRESSURE 940 mbar
FB HP T° 88°C
1
- Number of running hours
2
- LOC for local operating mode
(
C 10
)
3
- N2 for Purge ON
4
- Roots ON
5
- Display of pump information*
1
2
3
4
The user and /or OEM are ultimately responsible for operating the equipment
in a safe manner. The manufacturer has no control over the types of gases
exposed to this pump. This is the user and/or the OEM’s responsibility to
follow the necessary safety requirements.
Frequently process gases are toxic, flammable, corrosive, explosive and/or
otherwise reactive.
Toxic gases can cause serious injury or death. Operators and users must
take the appropriate safety recommendations to prevent injury. Consult the
responsible department for instructions and safety information.
5
adixen Vacuum Products - A3H Series Operating instructions