G 03898 - Edition05 - August 1
1/4
C 10
Operating mode
In local mode, the remote control cover plug must be connected to
the remote control connector at the rear panel of the pump.
The remote cover plug is necessary to desable the remote control mode and
to enable the local control mode.
The pump is running as a stand-alone part of the equipment on
which it has been installed.
The pump is controlled by the hand held remote (HHR), connected
in front or rear panel of the pump.
Switch configuration
according to the HHR
layout
Different
control modes
Local control mode
According to the monitoring configuration and wiring, the pump
can be:
locally controlled by the hand-held remote (HHR),
remote controlled by the equipment (remote control connector),
remote controlled by serial link (RS232 connector)
Two switches have to be configurated: the first at the back of
HHR, near the cable connector, and the second,
S2
, located on M4
electrical interface (
A 20
).
Even if 2 HHR can be connected together, one in front and one at the rear,
for safety reasons, the pump can be controlled only by one HHR at the same
time.
If the EMO connector is not connected to the tool, it is necessary to connect
the EMO cover to the EMO (J1) connector
B 100.
1
S1
J4
2
ON
1 2 3 4
ON
Connected at the rear
1
S1
2
ON
1 2 3 4
ON
Connected in front
Use of two hand-held remote control :
1
S1
2
ON
1 2 3 4
ON
2
1
3 4
ON
1
S1
J4
2
ON
1 2 3 4
ON
1 2 3 4
ON
Connected on both sides
Connected in series
Use of the hand-held remote control :
adixen Vacuum Products - A3H Series Operating instructions