G 03908 - Edition 03 - Februar
y 10
1/1
C 60
Use of the “Power failure protection”
The mains voltage
is restored within a
delay of 1 seconds
The mains voltage
is not restored
within the second
following the mains
power failure
This device includes a board installed in the electronic cabinet which
supplies power to the electronics for a second in the event of main
power failure.
The option can be configured by the user in the hand held remote
control DEFINITION menu.
Select the battery “ENABLED” option in the hand held remote
control DEFINITION menu. If a mains power cut occurs, a message
“W09 power supply”
is displayed on the hand held remote
control,
but the state of output contacts don’t change on the
remote control connector
.
Two cases can arise :
If the pump is running when the failure occurs, it will restart
automatically and goes back to initial operation.
After this time a message
“D09 power supply”
is created (as the
pump is no longer powered, this message will be displayed once
the mains supply will be restored, in the latest alarms section of the
ADMINISTRATION menu).
The battery supply is then cut.
This alarm will thus be taken into account by customer’s equipment.
When the main voltage is restored:
– if
“AUTO RESTART”
is DISABLED, the pump stands stopped,
– if
“AUTO RESTART”
is “ENABLED”, the pump restart
automatically ond goes back to initial operation.
adixen Vacuum Products - A3H Series Operating instructions