C 44
Saving and loading of pump
configuration (with HHR)
G 03905 - Edition 03 - February 10
Remote loading
Connect the M4 to the pump to configure.
Take the control of the pump from the HHR (blinking of “*” and
“X” at left corner of the display).
Load the configuration previously saved into the pump by selecting
the
CONFIG. LOAD
menu from MANAGEMENT menu.
The former configuration is automatically reset.
All pump parameters are loaded remotely, except the options which
have to be locally configurated according to the pump:
language
buzzer
serial link
date and time
temperature unit
pressure unit.
Loading parameters may result in creating software conflicts.
After loading, configure manually the following parameters (temperature
unit, pressure unit, serial link, time and date).
adixen Vacuum Products - A3H Series Operating instructions