G 03887- Edition 04 - August 12
1/2
B 20
Positioning the pump
in the pumping installation
Positioning the pump
Determine where the pump will be placed.
Refer to the technical specification section for dimensions, if needed
(see
A 40
).
Use the handling devices to position the pump in the desired
location (see
B 10
).
Pump performance will depend on the kind of accessories used and the
quality of the mechanical connections such as the pump fittings.
As these pumps are typically used in a corrosive atmosphere, their reliability
will depend on proper installation and maintenance. When assembling the
vacuum circuit, we recommend the installation of maintenance accessories
such as shut-off valves on the inlet and purge lines.
For safety reasons, use accessories on the inlet and exhaust lines whose
materials and sealing properties are compatible with the gases being used.
The pump must be operated in the horizontal position in support on its feet,
with the pumping axis vertical and the inlet opening upwards.
Lock the pump by adjusting the for
leveling feet for resting solidly on
the floor.
Use a water level to check if the
frame is in horizontal position in
both axis.
Risk of tilting
Although the appliance meets EEC safety regulations (topple angle ± 10°),
it is advisable to guard against the risk of tilting during handling,
installation, and use.
Refer to A40 of the user’s manual for location of the center of gravity.
Don’t place the product on a tilted surfaces, it can drag the operator
along due to its heavy weight:
use castors to move it on short distances,
locate it on flat, hard ground,
do not push it sideways
adjust the levelling feet for operation to make sure that the pump is
not supported by the castors.
The product is not designed to carry people or loads and is not for use
as a seat or step.
–
–
–
–
Risk of trips.
Route and secure cables, hoses and pipework during installation to
avoid any possible risk of trips.
adixen Vacuum Products - A3H Series Operating instructions