4/5
C 10
GB 02886 - Edition 03 - Januar
y 09
Operating modes
To change the parameter of the
Standby N
2
option
from
Disabled
to
Enabled
with the HHDM, perform the following operations:
Connect the HHDM plug to the HHDM connector on the front or rear
panel.
Set the
Standby N
2
option
parameter in the
DEFINITION
menu
C 42
as follows:
Standby N
2
option
Enabled
Hand held display module
message
A message specific to each stand-by purge configuration will be
displayed on the HHDM.
HHDM
message
Setting of the monitoring
Equipment command
Standby N
2
option
Via remote control connector (S5)
N
2
Disabled
/
* *
Enabled
Activated
N *
Enabled
Not activated
Autostart control
operation
The
autostart function
allows the automatic restarting of the pump when
the stop is caused by a power failure.
«Auto-start» setting
The selection of the
Auto-start
can be performed by the RS 232 / 485
or the HHDM (accessory).
To change the parameter of the
Auto-start
from
Disabled
to
Enabled
with
the HHDM, perform the following operations:
Connect the HHDM plug to the HHDM connector
Set the
Auto-start
parameter in the
DEFINITION
menu
C 42
as
follows:
Auto-start
Enabled
When the pump has been configured for automatic restart after a power
failure, it is the responsibility of the user to take all the measures required to
prevent risks resulting from this type of operation.
Stand-by purge
operation (option)
The
stand-by purge function
can be used to reduce the pump
nitrogen consumption during process idle. The «remote control» plug
B 70
must be used for this feature.
«Standby N
2
option» setting
The selection of the
Standby N
2
option
can be performed by the
RS 232 / 485 or the HHDM (accessory).
adixen Vacuum Products - Operating instructions - A3P Series