GB 02886 - Edition 03 - Januar
y 09
1/5
C 10
Operating modes
Different control
modes
According to the monitoring configuration and wiring, the pump can
be:
Locally controlled by the hand held display module (accessory).
Remote controlled by the equipment (remote control connector).
Remote controlled by the serial link (RS 232 connector).
Local control mode
Local control with hand held
display module (accessory)
The pump is running as a stand-alone part of the equipment on which
it has been installed.
The pump is controlled by the hand held display module (HHDM
accessory), connected in front or rear panel of the pump, or by both
simultaneously connected, one in front and the other at the rear panel
of the pump.
A second HHDM can also be connected in series with the first one.
Remote control modes
Remote control by «J14
remote control connector»
Pump alone
The pump is monitored by the equipment in which it is installed (such as
an automated pumping system or industrial equipment).
The pump is remote controlled by the opening or closing of different dry
contacts wired on remote control connector
B 70
.
This mode is enabled when
S1
contact is open that also inhibit the
control of the pump by the HHDM. However, the data and operating
parameters can be read on the HHDM.
In remote mode, «
REM
» is displayed on the HHDM.
Remote control via the
«RS 232 connector»
Pump alone or in a group of pumps
The pump is monitored by the equipment in which it is installed, either
alone or in a group of pumps.
The pump is remote controlled by the commands transmitted on the
serial link (refer to
B 110
for wiring and
C 90
for commands).
In local mode, the remote control cover plug must be connected to the
remote control connector at the rear panel of the pump.
The remote cover plug is necessary to desable the remote control mode and
to enable the local control mode.
For safety reasons, the pump can be controlled only by one hand held display
module at the same time.
If the EMO connector is not connected to the tool, it is necessary to connect
the EMO cover to the EMO connector
B 100
.