GB 02874 - Edition 02 - November 07
1/1
B 20
Positioning the pump in the pumping
installation
Positioning the pump
Determine where the pump will be placed. Refer to the technical
specification section for dimensions, if needed
A 40
.
Use the handling devices (accessory) to position the pump in the desired
location
B 120
.
(/2):/.4!,352&!#%
%XHAUST
02/()")4%$
)NLET
Each pump is equipped with four locking screw jacks. Lock the pump
by adjusting these jacks so that all four feet are resting solidly on the
floor.
,OCKINGSCREWJACKS
Pump performance will depend on the kind of accessories used and the
quality of the mechanical connections such as the pump fittings.
A3-P Range pumps are used in corrosive environments. Their reliability
will depend on proper installation and maintenance. When assembling the
vacuum circuit, we recommend installing maintenance accessories such as
isolation valve on the inlet, exhaust and purge lines.
For safety reasons, the materials and sealing properties of the accessories
used on the inlet and exhaust lines must be compatible with the gases being
used.
The pump must operate in the horizontal position. It may not be used
standing on its end.
Risk of tilting: compliance with the EEC safety rules is guaranteed (normal
range ±10°). However, precautions should be taken against the risk of
tilting during product handling, installation and operation. See
A 40
for
location of the center of gravity.
adixen Vacuum Products - Operating instructions - A3P Series