GB 03063 - Edition 04 - Oct 13
1/3
B 51
Checking the direction of rotation at initial
pump start-up (for A603P pump only)
Preliminary operations
Switch on the pump
Position the main switch on position «
1
».
Main switch
Adixen Vacuum Products
MONITORING VX.XX
PRESS ENTER
OR AUTO CONNECTION
The hand held display module allows to control
the pump and to configurate the parameters.
Connect the hand held display module.
FOR ALL PUMP
SERIAL XXXXXXX
At the first hand held display module connection,
the monitoring identifies the pump.
The display lights up and the following screens
appear:
MAITING .....
OLWRTA
Initialize the system by pressing on
ENTER
.
Controlling the pump using
the hand held display
module
23 H LOC N2 LS RT
PUMPING STOPPED
15/10/05 10H10
The monitoring displays the status of the
pump.
Because it is the first connection, it is necessary
to take the pump control via the
CONTROL
key:
a star appears on the left side of the display
.
A603P pump model is not equipped with a frequency converter:
the motor rotation depends directly on the main power connection.
Uncorrect wiring may cause backwards pump rotating.
Its is necessary to check the direction of rotation at first start up.