Operation monitoring
During operation, the user is warned of an operating incident by:
The audible buzzer warning (if preselected
C 42
).
Indicator lights on the front panel are illuminated.
Green on
: pump is running and no problems
Yellow and green on
: warning
Red on
: hazard
'REENLIGHT
9ELLOWLIGHT
2EDLIGHT
&RONTPANEL
Pumping is stopped when an hazard threshold is reached (if option
selected).
The fault contacts on the remote control connector at the rear side of
the pump are open.
A message via the RS 232 serial link.
If the pump is equipped with hand held display module (accessory):
One or several faults are displayed on the screen, alternately
with monitoring parameters. When these faults are activated, the
monitoring triggers the warning phase which is followed by the
hazard phase. The period of time for these phases can be set
C 42
. Only installed and configured parameters are displayed.
Indicator lights on the hand held display module are illuminated.
Study the safety instructions related to maintenance
D 00
.
adixen Vacuum Products - Operating instructions - A3P Series