Electrical connection
Electrical hard wire
connection
2EARPANEL
-AINPOWER
2EARPANEL
!!PUMPS
(ARDWIRE
FEEDTHROUGHT
!PUMP
Without Semi option
Unlock the black handles.
Loosen (1/4 turn) the two blocking screws of the left upper panel
(exhaust side view).
Open the panel.
Pass the cable through the feed through provided for this purpose.
Connect the mains cable to the main switch using a cable with following
specification:
Cable (customer supply)
A103P
AWG-14 (or 2.08 mm
2
)
A603P / A1003P
AWG-10 (or 5.26 mm
2
)
To ground the unit use a ring lug on the ground conductor (green /
yellow) and attach to the electrical frame pin with M4 nut.
Replace the left upper panel.
2INGLUG
-AINSWITCH
Check that the cable is correctly protected against ground defects and that
earth wire is longer than the three conducting wires.