Safety instruction related to maintenance
When the EMO is activated, some internal electrical parts are energized and
water cooling line and N
2
purge line are always supplied.
Risk of electrical shock in case of contact.
Before any intervention, disconnect main electrical cable, water inlet and
outlet quick connectors and N
2
purge connector.
Hazardous voltage enclosed.
Voltage or current hazard sufficient to cause shock.
Disconnect and lockout power before servicing.
Any intervention must be done by trained personnel only.
When the pump is switched off, internal parts (monitoring, frequency
converter) contain capacitors charged with over 60 V DC and remain
energized.
Electrical shock may result in severe injury.
Wait 1 minute after switching off before opening the pump.
Located on the upper cover, this label warns the user against possible
risk of injury due to any hand contact with hot surfaces. It states that
protective gloves should be used before performing any intervention.
The protective gloves should be used in accordance to the protective
gloves supplier’s instruction.
Located on the upper cover, this label indicates that some of the
internal parts are energized and could cause electrical shocks in case
of contact. It advizes to disconnect the pump before any intervention
or to properly lock-out and tag-out the equipment breaker before any
intervention on the pump.
A 329 982 - 2 and A 215 138 - 2
A 329 982 - 3
Manufacturer contact
The partial and full overhaul must be performed by Manufacturer
trained personnel.
Contact Manufacturer nearest service center or the service support at
the following e-mail address:
[email protected]
When the main switch is set to «0», the part supply power and the main
switch is energized.
Risk of electrical shock in case of contact.
Disconnect main electrical cable before any intervention.
This chapter describes the main preventive maintenance operations and
provides a guide for operating the equipment safely.
adixen Vacuum Products - Operating instructions - A3P Series