8/8
D 30
Diagnosis and trouble shooting
GB 04155 - Edition 04 - Oct 13
INCIDENT
CAUSE
TROUBLESHOOTING
Safety hazard
D44 SAF. CABINET PRES
Pump not equipped with Semi
option
Strap disconnected
Check the exhaustability
connection on the «Enclos.
switch» connector
D 10
.
Other problem
Call the customer service center.
Pump equipped with Semi option
No exhaustability in the Exhaust
box
Check the customer exhaustability
line.
Exhaustability threshold
< 1.6 hPa
Check the customer exhaustability
line.
Sensor problem
Check the exhaustability
connection on the «Enclos.
switch» connector
D 10
.
Other problem
Call the customer service center.
Safety hazard
W45 START UNABLED
Safety or Maintenance plugs
disconnected
Connect the safety or maintenance
plugs.
Other problem
Call the customer service center.
Hazard
D62 ROOTS FREQ. CONV.
Call the customer service center.
adixen Vacuum Products - Operating instructions - A3P Series