4/6
C 90
GB 02895 - Edition 02 - Oct 13
Use of the serial link
Order
Paramater
Response
Action
TPS
XX
Maintenance time XX =
00 :
running
time
01 :
full
overhaul
02
:
P out maintenance
03 :
reserved
04 :
exhaust
Maintenance
05 :
Reserved
06 :
Reserved
ROO
ON
or
OFF
OK
or
ERR2
Start up the ROOTS with the command ROOON and stop with
ROOOFF.
PUR
ON
or
OFF
OK
or
ERR2
Start up the purge with PURON, and stop with PUROFF.
DEF
XXY
FAULT
Display the 10 last Alerts or Alarms
Example :
# 0 0 0 D E F X X Y
XX =
0 to 09
Y =
0 alarm
1 alert
STA
None
Transmission of a Data Logger frame
Example :
#000 00100010000 0990 0 000 017 000 000 000 00 000000000000000000000 0000
0/3 5/15 17/20 22/25 27/29 31/33 35/37 39/42 44/45 47/67 69/72
0
:
header character
1-2-3 :
address
4
:
separator character
5 :
pump
status
6
:
roots status
7 :
purge
status
8
:
Stanby control mode
bit 0 = 0 :
stand-by purge Off
bit 0 = 1 :
stand-by purge On
bit 1 = 0 :
stand-by power Off
bit 1 = 1 :
stand-by power On
9 :
water
valve
10 :
inlet
valve
11 :
reserved
12
:
pump water valve
13 :
reserved
14 :
reserved
15
:
control in bit 3
and monitoring 0-1-2
16 :
separator
character
17-20 : pressure
(hPa)
21 :
separator
character
22-25 : power
(W)
26 :
separator
character
27-29
:
pump temp. (°C)
30 :
separator
character
31-33 : mfs
34 :
separator
character
35-37 : reserved
38 :
separator
character
39-42
:
analog input (mV)
43 :
separator
character
44-45
:
value hazard FC
46 :
separator
character
47 :
pressure
fault
48 :
reserved
49 :
reserved
50 :
reserved
51 :
reserved
52 :
reserved
53 :
frequency
control
54 :
reserved
55 :
reserved
56
:
motor temperature fault
57 :
reserved
58 :
reserved
59 :
N
2
purge fault
60 :
maintenance
fault
61
:
pump temp. fault
62 :
N
2
pressure fault
63 :
reserved
64 :
reserved
65 :
reserved
66 :
reserved
67 :
reserved
68 :
separator
character
69-72 : N
2
purging on second
73 :
CR
List of controls (continued)
adixen Vacuum Products - Operating instructions - A3P Series