Diagnosis and trouble shooting
INCIDENT
CAUSE
TROUBLESHOOTING
Warning
W08 WATER FLOW
No water flow
Check the water supply.
Insufficent water flow (< 60 l/h)
Check that the pipes are not
clogged.
Check that there are no leaks.
Sensor problem
Check the sensor connection on
the «Water switch» connector
D 10
.
Other problem
Call the customer service center.
Warning
W10 MOTOR TEMP.
Motor temperature is > 120 °C
Refer to water flow defect
(
see W08
).
Sensor problem
Check the sensor connection on
the «Motor temp. ADP switch»
connector
D 10
.
Other problem
Call the customer service center.
Safety hazard
D10 MOTOR TEMP.
or
Safety hazard
D41 SAFETY MOTOR TEMP.
D41 defect appears only if the
Sa-
fety plug
is connected to the safety
connector
B 80
Motor temperature is > 150 °C
Refer to water flow defect
(
see W08
)
Sensor problem
Check the sensor connection on
the «Motor temp. ADP switch»
connector
D 10
.
Other problem
Call the customer service center
Warning
W11 INLET VALVE
There is no isolating valve at inlet
Incorrect monitoring
configuration
Check that «INLET VALVE
OPTION» is disabled
C 42
.
Other problem
Call the customer service center.
There is an isolating valve at inlet
The valve does not close at pump
stop or does not open at start-
up
Check compressed air supply.
Check valve condition.
Other problem
Call the customer service center.