Maintenance
198
Connecting the Nebulizer Argon Tubing (Cross-Flow End Cap,
N0770546
)
1. Pass the tubing (Part No. 09985723) through the nut (Part No. 09920545) and
attach the ferrule (Part No. 09920515) on the end of the tubing. See the following
figure.
2. Insert the ferrule with the tubing and nut to the Male Quick Disconnect (Part No.
N0775206). Connect the entire Quick Disconnect to the sample introduction
system. See the following figure.
Replacing the Nebulizer Argon Tubing (Cross-Flow End Cap,
N0770546
)
If you replace the argon tubing with new tubing, you will need to replace the
Swagelok nut as follows. See the previous figure.
1. Using a razor blade, make a straight cut on one end of the tubing. Wire cutters
or scissors do not make acceptable cuts.
2. Put the 1/8-in. Swagelok nut (Part No. 09903128), rear ferrule, front ferrule, and
tubing insert onto the tubing as shown in the figure below.
3. Finger-tighten the nut on the Brass Reducer Fitting.
4. Connect the nebulizer argon tubing to the NEB connection. Using a wrench,
tighten the 1/8-in. nut an additional 3/4 turn.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......