System Description
65
Spectrometer
Introduction
The core of the optical system comprises a dual Echelle monochromator with a dual,
backside-illuminated, cooled, CCD detector. The system is specifically designed for
ICP-OES. Computer controlled transfer optics direct the radiation from the plasma
into the monochromator. The optics housing is sealed and continuously purged with
high purity nitrogen.
Signals at the required analytical wavelengths are measured using a scanning CCD
(charged coupled device) based technology, with simultaneous measurement of the
background emission and a neon spectrum for active wavelength correction.
Transfer Optics
The computer controlled transfer optics are used to select either radial or axial
viewing of the plasma and to direct the radiation from the plasma onto the entrance
slit of the monochromator. The exact viewing position, horizontal and vertical
position, can be selected in the software.
An automatic shutter closes between measurement cycles to reduce the exposure of
the optics to excess UV radiation.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......