Maintenance
175
Adjusting the Position of the Shear Gas Nozzle
The basic shear gas nozzle position is with a 1 to 2 mm gap between the nozzle slit
and the front edge of the axial window mount. For some special analyses you may
want to move the nozzle, for example, for samples containing organics or salts, set
the nozzle slit 4 mm in front of the front edge of the axial window mount.
1. Loosen the two Pozi Drive Head screws that hold the shear gas nozzle assembly
in position.
2. Slide the assembly to the position you require, then tighten the screws.
Caution
Risk of heat damage to the axial window.
Never
set the nozzle slit behind
the front edge of the axial window mount.